K914 covers testing of the electrical receptacles at patient bed locations. It is the second most-cited electrical tag in nursing homes after K920, and it is almost always cited as widespread: 72% of K914 citations were rated scope F. That happens because receptacle testing is usually all-or-nothing. If there is no record, every bed location is out of compliance.
Source: CMS Provider Data Catalog, Nursing Homes, Fire Safety Deficiencies and Provider Information (processed October 1, 2026), Life Safety Code surveys from October 2023 through September 2026. Scope and severity: 13% D (isolated), 12% E (pattern), 72% F (widespread).
What K914 requires
NFPA 99 requires receptacles at patient bed locations to be tested after initial installation, replacement or servicing, and the facility needs a written interval for ongoing testing. Each test checks four things:
| Test | What it checks |
|---|---|
| Physical integrity | Cracked faces, loose mounting, damaged contacts |
| Polarity | Hot and neutral wired correctly |
| Ground continuity | The grounding path is intact |
| Retention force | Each blade holds the plug firmly; NFPA 99 sets a minimum of 115 g (4 oz) per blade |
Surveyors cite K914 when the facility can’t produce a record of these tests by room and bed location, when the record is incomplete, or when they find receptacles that are visibly damaged or loose.
Where K914 is cited most
The highest K914 rates were in Maryland (40.3%), Maine (37.2%), Washington (32.1%), Hawaii (30.2%), Nevada (28.8%). Compare your state in the electrical citations report.
How to correct a K914 citation
- Test every receptacle at every bed location for integrity, polarity, ground continuity and retention, and log the result by room and bed.
- Replace failed receptacles (hospital-grade at bed locations) and re-test them. Log the replacement and the re-test.
- Set the interval in writing (many facilities test annually) and assign who owns it.
- Keep the log with your PCREE records. Surveyors often ask for K914 and K921 documentation in the same request.
The median facility corrected K914 in 37 days. Receptacle testing and PCREE testing use the same kind of electrical safety analyzer, so many facilities schedule both in one visit. See receptacle testing in patient rooms.
Free: Electrical K-Tag Plan of Correction Worksheet
Covers K911 through K921, including K914, with a revisit evidence checklist and typical correction times by tag.
Instant download. We never share or sell your information.
Get a Free Quote Draft a K914 Plan of Correction Look Up Your Facility