K911, K912 and K919 are the less common electrical Life Safety Code tags, but they are rarely cited alone. They show up when a fire safety surveyor finds a problem with the building’s wiring, receptacles or how equipment is used, and they usually lead to closer review of K914, K920 and K921 on the same survey.

TagWhat it coversFacilities cited (3 yrs)Median days to correct
K911Installation and maintenance of electrical systems (panels, wiring, GFCI, cover plates)911 (6.2%)41
K912Properly grounded receptacles, including hospital-grade receptacles at patient bed locations546 (3.7%)34
K919Use of electrical equipment, including resident-owned appliances659 (4.5%)26

Source: CMS Provider Data Catalog, Nursing Homes, Fire Safety Deficiencies and Provider Information (processed October 1, 2026), Life Safety Code surveys from October 2023 through September 2026.

K911: electrical systems

Common findings: missing cover plates, open junction boxes, panels blocked by storage, unlabeled circuits, and GFCI receptacles that don’t trip. Fixes are usually quick, but they need a licensed electrician and a dated record. Highest rates: Montana (42.6%), New Mexico (32.4%), Maine (24.4%), New Jersey (20.7%), Wisconsin (20.1%).

K912: receptacle grounding

Common findings: receptacles that fail a ground test, worn or cracked receptacles at bed locations, and non-hospital-grade receptacles where hospital grade is required. K912 overlaps with K914 receptacle testing: a facility with current K914 records usually finds K912 problems before the surveyor does. Highest rates: Wisconsin (25.7%), Montana (19.7%), Pennsylvania (18.1%), Nevada (10.6%).

K919: use of electrical equipment

Common findings: resident-owned electric blankets, heating pads or appliances in use without an inspection, damaged cords and plugs, and adapters that defeat the ground pin. 15.6% of facilities cited for K919 were cited again later, so admission checks of resident-owned items matter. Highest rates: California (28.3%), Maine (23.1%), Nebraska (18.9%), Alaska (15.0%), Connecticut (14.7%).

How to correct these citations

  1. Fix the specific findings with a licensed electrician (K911, K912) or by removing and replacing the item (K919), and keep dated records.
  2. Test receptacles at every bed location if you don’t have current records (K914).
  3. Confirm every patient care device has current, documented PCREE testing (K921). Surveyors who cite one electrical tag usually check the rest.
  4. Write or update the policy for resident-owned electrical items and train admissions and nursing staff.

Free: Electrical K-Tag Plan of Correction Worksheet

A worksheet for K911 through K921 with a revisit evidence checklist and typical correction times by tag.

Instant download. We never share or sell your information.

Get a Free Quote Draft a K911 Plan of Correction Draft a K912 Plan of Correction

Frequently Asked Questions

What is a K911 citation?
K911 is the CMS Life Safety Code tag for the installation and maintenance of electrical systems in health care occupancies, such as panels, wiring, cover plates and GFCI protection.
What is a K912 citation?
K912 is the tag for properly grounded power receptacles, including hospital-grade receptacles at patient bed locations.
What is a K919 citation?
K919 covers the use of electrical equipment, including resident-owned appliances and devices with damaged cords, plugs or adapters.
Do these citations mean we also need PCREE testing?
Often, yes. A surveyor who finds an electrical problem usually asks for K914 receptacle records and K921 PCREE testing records next.