Walk through almost any skilled nursing facility and you'll find power strips: behind a resident's recliner, powering a TV and phone charger, or under a bed next to an air mattress pump. They're convenient, and they're also one of the most common things a Life Safety Code surveyor notices on a walk-through.

By the numbers: Power strips and extension cords (K920) are the most-cited electrical Life Safety Code tag in U.S. nursing homes: 5,299 citations at 4,170 facilities from September 2023 to August 2026. California alone accounted for 1,100. See the state-by-state data.

Why Power Strips Get Facilities in Trouble

A power strip (the codes call it a relocatable power tap, or RPT) multiplies what can be plugged into one receptacle. In a resident room that can mean medical equipment and personal electronics sharing an overloaded strip, strips plugged into other strips, or a cheap consumer strip powering an air mattress. Each of these is a fire and shock risk in a building full of people who may not be able to evacuate on their own.

The CMS Power Strip Waiver

For years, the older code edition CMS enforced made power strips in patient care areas very hard to use. In 2014, CMS issued a categorical waiver (Survey & Certification letter S&C 14-46-LSC) that lets facilities use power strips in patient care rooms if they follow the newer NFPA 99 (2012) requirements. As summarized by the American Society for Health Care Engineering, the key conditions are:

  • Medical equipment powered from a strip must use a special-purpose relocatable power tap listed to UL 1363A or UL 60601-1, not a hardware-store strip.
  • Non-medical equipment (TVs, lamps, chargers) may use a power strip listed to UL 1363.
  • The facility must have written documentation that it has elected to use the waiver.
  • Staff must tell the surveyors at the entrance conference that the facility is using the waiver and meets its conditions.

Since 2016 CMS has enforced the 2012 editions of NFPA 99 and NFPA 101 directly. Many facilities still keep their waiver election on file, and surveyors still check the underlying conditions. Confirm the current expectations with your state survey agency or Life Safety Code consultant.

Practical Rules for Resident Rooms

  • Never mix medical equipment and personal electronics on the same consumer power strip.
  • No daisy-chaining: a power strip should never be plugged into another strip or an extension cord.
  • Keep strips off the floor where they can be rolled over, kicked or soaked, and out from under bedding.
  • Don't use strips for high-draw items such as space heaters, refrigerators or microwaves.
  • Inspect strips and cords on rounds, and replace any with damaged housings, scorch marks or loose receptacles.
  • Check resident-owned items at admission. See resident-owned equipment and PCREE.

How This Connects to PCREE Testing

Power strips that are part of a medical equipment setup (for example, a strip mounted to a cart or bed system that powers several devices) are part of that system. NFPA 99 expects a system of several electrical appliances to be compliant as a complete system. That means your PCREE technician should test it as installed, not just the individual devices. When you schedule testing, tell the technician where medical-grade strips are in use so they're included in the scope and listed on the report.

Want power strips and equipment assemblies included in your next PCREE test?

Get a Free PCREE Quote →

Free: PCREE Survey-Readiness Checklist

The one-page list surveyors work from: inventory, test records, failures and re-tests, technician documentation.

Download the Checklist →

Free: Power Strip & Extension Cord Audit Sheet

A room-by-room walk-through sheet for K920, the most-cited electrical tag in nursing homes.

Instant download. We never share or sell your information.

Frequently Asked Questions

Are power strips allowed in nursing home resident rooms?
Yes, with conditions. Under the CMS power strip waiver and NFPA 99 (2012), power strips used with medical equipment must be special-purpose units listed to UL 1363A or UL 60601-1, power strips for non-medical items must be listed to UL 1363, and the facility should document its use of the waiver and tell surveyors at the entrance conference.
Can a resident's TV and an air mattress pump share a power strip?
They should not share a consumer power strip. Medical equipment such as an air mattress pump needs a medical-grade, UL 1363A or UL 60601-1 listed power tap or its own receptacle. Personal electronics can use a UL 1363 listed strip.
Does PCREE testing include power strips?
Medical-grade power strips that are part of an equipment assembly should be tested with that assembly, because NFPA 99 expects a system of several appliances to be compliant as a whole. Tell your technician where they are so they are included in the scope.

About the Author

PCREE Test Editorial Team

Content reviewed by biomedical professionals with experience in patient care electrical equipment testing, NFPA 99 compliance, and CMS Life Safety survey preparation for skilled nursing facilities.